Example of installation with PVC mounting
bracket and BAMOWIZ
Optical fiber IR measurement
From 5 to 4000 NTU/FNU or 0 to 4500 mg/L
90° measurement using infrared at 850 nm
Modbus RS-485 digital output
In accordance with standard 7067
APPLICATIONS
- Urban wastewater treatment (inlet/outlet controls) - Sewerage network (load monitoring) - Industrial water treatment - Surface water monitoring, dredging site
DESCRIPTION
The TURBINUM sensor measures turbidity using the nephelometric principle. It operates with a diode that emits infrared light at a wavelength of 850 nm, and another diode positioned at a 90-degree angle that detects the scattered radiation.
This optical technology requires minimal maintenance. The probe stores its calibration data internally. All measurements are processed directly within the sensor and transmitted via MODBUS communication.
The BAMOWIZ TUR, specifically developed for the TURBINUM sensor, provides local display and signal management. The device is equipped with a touchscreen for displaying measurements and includes additional functions such as alarm threshold configuration, 4-20 mA signal retransmission, and sensor calibration. It also features two 4-20 mA inputs for additional measurement capabilities. (See datasheet 442-01)
General Catalog 442-02 17/07/2025 16:49
Digital Turbidity Sensor TURBINUM 442
BAMO Mesures SAS
22, Rue de la Voie des Bans Parc d'activités de la Gare
95100 ARGENTEUIL France