Recherche
MENU Catalog Global Sites
BAMO
INTERNATIONAL
European manufacturer of sensors and instruments for liquids
You are in BAMO EU » General Catalog » Chlorine, disinfection » CCK
General Catalog 193-12
CCK
CCK
Chlorine dioxide and free chlorine sensor
CCK

  • For industrial hot water, drinkable water, cooling water
  • Range: from 0.01 up to 4 ppm
  • Potentiostatic electrode
  • Reversible Voltage for automated cleaning operation
  • Pressure limit: 6 bar as a maximum
  • Unnecessary zero adjustment
DESCRIPTION

Principle
Potentiometric method with a cell of 3 electrodes; without temperature sensor. A Pt 100 sensor must be put in
place for compensation signal.

Automatic cleaning system
By reverse polarity included inside regulator.

Mounting / Recommendations
The measuring at a constant flow rate requires the use of a specific cell
holder (see data sheet 193-95).
The complete assembly optimizes the operations.

TECHNICAL FEATURES

Range: 0.01 to 4 ppm
pH range: 5 to 9
Operating pressure: 6 bar as a maximum
Operating temperature: 5 to 70°C
Flow rate limits: 30 to 40 L/h (see data sheet 193-95)
Mounting: Thread PG13.5
Connection: Specific coaxial cable (5 m long) M12 connector
Materials: Glass, gold electrodes
Dimensions: Ø 12 mm, length 120 mm

General Catalog 193-12
09/04/2018 17:28
Chlorine dioxide and free chlorine sensor
CCK
Technical & Commercial support's main contacts
Ask a QUOTATION online
BAMO Mesures SAS
22, Rue de la Voie des Bans
Parc d'activités de la Gare
95100 ARGENTEUIL France
Tel : +33 (0)1 30 25 83 32
Fax : +33 (0)1 34 10 16 05
Email : export@bamo.fr
Site : http://www.bamo.eu
Siret : 314 055 864
TVA Intra : FR 08 314 055 864
APE : 2651 B
Capital : 400 000 €